| Model Number |
PCB00 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6b - explant date: not applicable, lens remains implanted, therefor not explanted.Section e1 - telephone number: (b)(6).Section h3 - other (81): the intraocular lens (iol) was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery day the customer had five cases of foreign material entering the eye during implantation.Previous cases had been reported by them and on this surgery day, the customer used healon pro instead of their normal (ophthalmic viscosurgical device) ovd to try and exclude the ovd as the source of the particle's they are seeing.Because the ¿extra material¿ was still observed, it would appear that the pre-loaded lens system seems to be the source.No further information was provided.Separate reports will be submitted for the other 4 incidents on the surgery day.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3: device evaluated by manufacturer - yes device evaluation: the customer provided a video showing an intraocular lens claimed to be a tecnis itec preloaded model pcb00 and a particle.Per the complaint description, the particle remains in the patient's eye and attached to the lens.The nature, root cause, and the potential clinical impact of the described particle cannot be determined from a video assessment.The complaint issue of foreign material - loose was identified during product evaluation; however, based on the complaint investigation results, the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: based on the complaint investigation results, the product was released within specifications.The complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|
|