• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NALU MEDICAL INC NALU NEUROSTIMULATION SYSTEM; SPINAL CORD STIMULATOR Back to Search Results
Model Number 71005
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 06/29/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Nalu rep was unable to attend the surgical procedure.All components, internal and external, were discarded at the time of the procedure and are thus not available for examination by nalu.There is no clear indication as to the cause of the system fluctuation in providing therapy.Migration and lead fracture were ruled out through xray imaging.
 
Event or Problem Description
Patient was implanted with the nalu spinal cord stimulator on (b)(6) 2021.Patient reported system began to function sporadically and not provide consistent therapy.Patient chose to follow up with a different doctor.Xray imaging was performed and did not indicate any obvious issues with the implanted components.New doctor recommended to explant the entire system and start fresh with all new internal and external components.System replacement was performed on (b)(6) 2023 and replaced the implantable pulse generator, leads, and external therapy discs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Common Device Name
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17414820
Report Number3015425075-2023-00166
Device Sequence Number21076705
Product Code GZB
UDI-Device Identifier00812537033600
UDI-Public01008125370336001121052617240526
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/26/2024
Device Model Number71005
Device Catalogue Number71005
Device Lot NumberA02293
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2023
Initial Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-