• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA BILIARY; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00539260
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific that a flexima biliary stent was attempted to be used during an endoscopic retrograde cholangiopancreatography procedure in the common bile duct, performed on (b)(6) 2023.During the procedure, the guidewire was stuck in the advancer.Another flexima biliary stent was used to successfully complete the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.This event has been deemed a reportable event based on the investigation results; guide catheter detached/separated.Please see block h10 for full investigation details.
 
Additional Manufacturer Narrative
Block e1: initial reporter address 1: (b)(6).Block h6: imdrf device code a0401 captures the reportable investigation result of guide catheter detached/separated.Block h10: the returned flexima biliary stent was analyzed, and a visual evaluation noted that the guide catheter was detached into two parts and was accordioned over the guidewire.The stent was not returned.No other problems with the device were noted.The reported event of guidewire stuck was confirmed.Based on all the gathered information, the device problems found suggests that the physician faced some adversities while introducing/removing the guidewire.The device problems may have been caused by the technique used or patient anatomy.Therefore, the most probable cause is adverse event related to the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA BILIARY
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17414955
Report Number3005099803-2023-03955
Device Sequence Number15249321
Product Code FGE
UDI-Device Identifier08714729162575
UDI-Public08714729162575
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K965147
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2025
Device Model NumberM00539260
Device Catalogue Number3926
Device Lot Number0029421386
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/03/2023
Initial Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight65 KG
-
-