Zimmer biomet complaint number (b)(4).Implant date has been updated to unknown.Zimvie received one (1) oss311, (osseotite® implant 3.75 x 11.5mm) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, apparent bone / tissue attached to external threads.The implant was fractured at the collar region.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number 2022060865.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022060865 for similar events and no other complaint was identified.Review completed utilizing keywords: fracture implant.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are missing or confusing instructions for use, clinician selects implant that is unable to resist long-term occlusal loading.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|