A one opened broken phaco tip, without a wrench, in a bag, with no lot info was received for the report.The phaco tip was visually inspected and deemed nonconforming, the phaco tip was broken in the middle, breakage is slightly jagged.The phaco had even wall thickness.There was wear on the threads and back of flange that was consistent with threading on a handpiece.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.The complaint evaluation confirms the phaco tip was broken.The root cause for the broken phaco tip cannot be determined from this evaluation.The phaco tip visual inspection does not show any manufacturing issue that would cause the broken phaco tip.No specific action with regard to this complaint was taken because the root cause for the complaint issue cannot be determined from this evaluation.All phaco tips are 100% visually inspected by trained operators using 30 x magnification during the manufacturing process.Any nonconformance, such as the broken phaco tip exhibited on the returned opened sample, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|