• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC VIPER PRIME INSERTER SHAFT; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER PRIME INSERTER SHAFT; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 286750031
Medical Device Problem Codes Device Slipped (1584); Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2023
Type of Reportable Event Malfunction
Event or Problem Description
Device report from germany reports an event as follows: it was reported that on (b)(6) 2023, the hcp was found to have problems with detaching the screw from the screwdriver.Sometimes the screw is hard to detach, and sometimes the screw sits a bit loose on the screwdriver tip.The screwdrivers seem worn.There was no reported patient impact, no reported surgical delay.No further information is available.This report is for a viper prime inserter shaft.This is report 6 of 7 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for viper prime inserter shaft, was conducted identifying that lot number mf4440501 was released in one batch.Batch1: lot qty of (b)(6) units were released on jul 26, 2021 with no discrepancies.Supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the viper prime inserter shaft.The overall device appears to be in good condition with no signs of wear of the tip that could have contributed to the functional issues reported.A dimensional inspection was performed for the viper prime inserter shaft and met specifications.A functional test was unable to be performed since its mating device was not returned.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the viper prime inserter shaft was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME INSERTER SHAFT
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17415518
Report Number1526439-2023-01393
Device Sequence Number17926016
Product Code NKB
UDI-Device Identifier10705034507583
UDI-Public(01)10705034507583
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K171570
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number286750031
Device Lot NumberMF4440501
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/12/2023
Supplement Date Received by Manufacturer09/05/2023
Initial Report FDA Received Date07/28/2023
Supplement Report FDA Received Date09/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN SCREWS; VERSE POLY DRIVER, SHAFT; VERSE POLY DRIVER, SHAFT; VIPER PRIME INSERTER SHAFT; VIPER PRIME INSERTER SHAFT; VIPER PRIME INSERTER SHAFT
Patient SexUnknown
-
-