SYNTHES GMBH 2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 30MM; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Catalog Number |
04.211.030S |
| Medical Device Problem Codes |
Break (1069); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from japan reports an event as follows: it was reported that on (b)(6) 2023, the patient underwent an orif surgery for distal humerus fracture.The affected area was fixed in the order of the medial plate, then the lateral plate.Initially, the screw was inserted at a fixed angle, but when it was checked with an image intensifier, it was found that the tip of the screw penetrated forward.The screw was removed and reinserted several times, but the angle did not change so the surgeon redrilled with a variable angle and inserted the screw.Near the end of the lateral plate fixation, the surgeon said that the 2nd screw from the distal side could not be locked and was coming out.The screw came out, so no screw was inserted in that screw hole.Only one screw could be inserted distally, so an additional screw was inserted from the distal lateral direction to the proximal lateral.The procedure was completed successfully with no surgical delay.No further information is available.This report is for a 2.7mm ti va lckng scr slf-tpng with t8 stardrive recess 30mm.This is report 2 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2: additional procode: hrs.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Part:04.211.030s.Lot:9l69007.Manufacturing site: werk selzach supplier:früh verpackungstechnik ag release to warehouse date:08 aug 2022.Expiration date: 01 aug 2032.Non-sterile part # 04.211.030.Non-sterile lot # 877p003.Manufacturing site: werk selzach supplier:synthes usa hq, inc release to warehouse date:14 jun 2022.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|