MICRO THERAPEUTICS, INC. DBA EV3 AXIUM PRIME BRPL HLX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
APB-1.5-3-HX-ES |
| Medical Device Problem Codes |
Break (1069); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that 3 coils had resistance/stuck in sheath.It was reported that there was a reoccurring carotid blowout intervention that required coiling.The university of michigan has been on a persistent backorder of axium coil consignment, and the manufacturer's representative had very little axium coils in their stock due to persistent backorders as well.The hcp wanted to use axium prime hx es coils in a 1.0mm and 1.5mm diameter size, but medtronic's persistent backorders prohibited the use of those specific sizes.The hcp had to settle for using axium prime 3d es coils instead through a medtronic echelon 14 microcatheter.Second qa issue occurred when the medtronic axium coils, which were all prepped according to the ifu, hydrating the coils in a bowl of saline then holding the coiling sheath in a vertical position retracted the coil back into the coiling sheath fully incapsulating the coil into the coiling sheath.During the coiling of the carotid branch segment in which a total of 14 axium coils were deployed into the carotid artery for embolization, four axium coils randomly felt by the hcp to have heavy resistance at the distal non-tapered segment of the coiling sheath and unable to insert the coil into the echelon 14 microcatheter.Some axium would deploy without issue while the other four axium coils randomly dispersed throughout the procedure would "get stuck" in the distal coiling sheath.The hcp noticed as the non-tapered axium coil sheath entered into the echelon 14 microcatheter hub a small gap(space) occurred between the distal coiling sheath and echelon 14 hub.As the coil traveled across that small gap of the non tapered coiling sheath the hcp could feel the high resistance across the gap and the coil would not advance any further.The defective axium coil would be removed and replaced with another which would randomly successfully deploy or get stuck once again.The reported devices and any accessory devices were prepared as indicated in the ifu.There was no patient involved in the event.Additional information was received that the hcp did report "feeling the coils break" at the gap between the coiling sheath and echelon microcatheter, but mostly removed the coils due to heavy resistance.Ancillary devices include a echelon 14 microcatheter: ref#: (b)(4).Lot#: b516198.
|
| |
|
Event or Problem Description
|
|
Additional information received reported there appeared to be no damage to the pushwire.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|