| Model Number |
BIO-COMP S-TAK 2.4X 12MM |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On(b)(6) 2023, it was reported by an arthrex employee via email that an ar-1934bcf-24-2 biocomposite suturetak anchor pulled out of the implantation site.This was discovered during a procedure on (b)(6) 2023.The case was completed using a new product.Additional information received on (b)(6) 2023: this was discovered during a labral repair procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misaligned insertion and/or prying/leveraging the device during insertion.
|
| |
|