| Model Number |
32375 |
| Medical Device Problem Codes |
Break (1069); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
Internal Organ Perforation (1987)
|
| Date of Event |
06/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient injury occurred.An 8.3 flexima drainage catheter was selected for right nephrostomy placement in the kidney.During the procedure, it was noted that the renal collecting system was ruptured due to inability to feed off catheter from metal stiffener.The patient has fully recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the stent was returned with the metal cannula for the analysis.Visual inspection was performed, and it was revealed that the hub and the suture were in good conditions.The metal cannula was kinked.The pouch was returned, and it was observed that the pouch information matches with the complaint information.No more issues were observed in the devices.
|
| |
|
Event or Problem Description
|
|
It was reported that patient injury occurred.An 8.3 flexima drainage catheter was selected for right nephrostomy placement in the kidney.During the procedure, it was noted that the renal collecting system was ruptured due to inability to feed off catheter from metal stiffener.The patient has fully recovered.
|
| |
|