| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 - report source - foreign: the event occurred in singapore.The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported during knee arthroplasty when the surgeon went to open the cementless femoral component, the packaging appeared dented, and the inner sterile packaging already partially opened.Due to the sterile barrier being compromised, the surgeon had to utilize a cemented femoral component to complete the procedure instead.No adverse events were reported as a result of this malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, g3, g6, h2, h3, h6, h10.Visual evaluation of the returned device identified that the inner and outer sterile blisters exhibited damage visually consistent with thermal damage.Further evaluation identified that the tyvek had lifted in multiple areas on both blisters.Sterility was confirmed to have been breached.The device history records were reviewed and no discrepancies were identified.The condition of the device when it left zimmer biomet was considered non-conforming to specification and therefore the root cause of the reported event is attributed to the packaging operator not following the work instructions provided.Corrective actions have been initiated to address the event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|