• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 5.0MM(D) X 1MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 5.0MM(D) X 1MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number ILPC541U
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number cmp(b)(4).A1: patient identifier unknown / not provided.A4: patient weight unknown / not provided.
 
Event or Problem Description
Doctor reported screw fracture inside of abutment head at tooth site 26.
 
Additional Manufacturer Narrative
Zimvie received one (1) ilpc541u, (certain low profile one-piece abutment 5.0mm(d) x 1mm(h)) for evaluation.Visual evaluation was performed, a fracture has been identified at the top.Dhr review was completed for the subject lot number 2018011908.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2018011908 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : screw.The customer did not submit images for the reported event.Based on the investigation and risk management file review as per rm-00057-haz, the most likely root cause determined from the investigation were missing or confusing instructions for use and clinician incorrectly engages abutment screw into implant.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the top.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H3: changed "no" to "yes".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN LOW PROFILE ONE-PIECE ABUTMENT 5.0MM(D) X 1MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key17423798
Report Number0001038806-2023-01431
Device Sequence Number7555914
Product Code NHA
UDI-Device Identifier00844868021032
UDI-Public(01)00844868021032(17)230219(10)2018011908
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K092341
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/18/2023
Device Catalogue NumberILPC541U
Device Lot Number2018011908
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/31/2023
Supplement Report FDA Received Date01/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/18/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-