• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RANGER PACLITAXEL-COATED PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RANGER PACLITAXEL-COATED PTA BALLOON CATHETER; DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER Back to Search Results
Model Number 1973-03
Medical Device Problem Codes Leak/Splash (1354); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the balloon ruptured and had a pinhole.This 4.0 x 60mm, 80cm ranger paclitaxel-coated pta balloon catheter was selected for use in an endovascular therapy procedure.The 99% stenosed, moderately calcified, and moderately tortuous lesion was located in the right superficial femoral artery.The balloon ruptured and exhibited a pinhole slightly distal from the balloon catheter center.This was noted during the first inflation at 8 atm for approximately five seconds.The balloon was removed intact, and the procedure was completed with another of the same device.The physician noted this was effects of calcification.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RANGER PACLITAXEL-COATED PTA BALLOON CATHETER
Common Device Name
DRUG-ELUTING PERIPHERAL TRANSLUMINAL ANGIOPLASTY CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
HEMOTEQ AG
adenauerstrasse 15
wuerselen 52146
GM   52146
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17424155
Report Number2124215-2023-40337
Device Sequence Number18056754
Product Code ONU
UDI-Device Identifier08714729975793
UDI-Public08714729975793
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P190019
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2023
Device Model Number1973-03
Device Catalogue Number1973-03
Device Lot Number05417H21
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2023
Initial Report FDA Received Date07/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-