CONMED UTICA BULB TIP YANKAUER W/VENT 50/CASE; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
|
Back to Search Results |
|
| Catalog Number |
0034880 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, 0034880, for an insufficient heatseal.The device was encroaching the seal.There was no contact with the patient as this was found during incoming inspection.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Received two 0034880 in original packaging.Lot number was verified.Performed a visual inspection, the device was encroaching the seal.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging did not have an insufficient heat seal on only one package while the second packaging had an insufficient heatseal.Root cause cannot be determined, however, based upon evaluation of the device; a possible cause of this event could be that the device was encroaching the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 8 reports, regarding 16 devices, for this device family and failure mode.(b)(4).We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|
Search Alerts/Recalls
|
|
|