BAUSCH + LOMB ENVISTA SIMPLIFEYE INJECTOR; FOLDERS AND INJECTORS, INTRAOCULAR LENS Back to Search Results
Model Number 21987
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/30/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on insertion of an intraocular lens (iol) into the patient¿s eye the optical plate of the lens was not intact.The lens was removed and another lens was implanted.The surgical time was extended and a suture was placed.Additional information was requested.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Additional information regarding the event was requested, but not received.The device history record (dhr) review did not find any non-conformities or anomalies related to this event.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause of this event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVISTA SIMPLIFEYE INJECTOR
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS
Manufacturer (Section D)
BAUSCH + LOMB
1400 n goodman st
rochester NY 14609
Manufacturer (Section G)
MEDICEL AG
dornierstrasse 11
alternhein CH-94 23
SZ   CH-9423
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key17425636
Report Number0001313525-2023-70096
Device Sequence Number7721799
Product Code MSS
UDI-Device Identifier00757770547449
UDI-Public(01)00757770547449(17)231231(10)2049A1
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K192005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number21987
Device Lot Number2049A1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENVISTA IOL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexUnknown
-
-