| Catalog Number |
19AGFN-756 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2023, a 19mm sjm regent mechanical heart valve, was selected for an implanted.At the time of device preparation, the leaflets was moving normally when tested with rubber hose.During the procedure, after implantation, the physician found that the valve leaflet did not open and close well.Device was explanted from the patient and a new 19mm sjm regent mechanical heart valve device was successfully implanted as a replacement.Post explant, the device was moving normally when tested.The patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of the valve leaflet did not open and close well after implantation was reported.Morphological and hydrodynamic examination indicated the valve met abbott specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Hydrodynamic testing upon return to abbott and at the time of manufacturing indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|