| Catalog Number |
115396 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Joint Dislocation (2374)
|
| Date of Event |
07/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: cat: 115330 lot: 289590 comp rvrs shdr glen bsplt +ha; cat: xl-115363 lot: 749930 arcom xl 44-36 std hmrl brng ; cat: 118001 lot: 047630 versa-dial/comp ti std taper; cat: 115370 lot: 284530 comp rvs tray co 44mm; unknown glenosphere.G2: foreign- eu: poland.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2023 - 01752.0001825034 - 2023 - 01766.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right shoulder was revised approximately seven months post-initial implantation due to the dislocation of the prosthesis.Intraoperatively, there was loosening of the central screw as well.Attempts to obtain additional information have been made; however, no more is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Devices are used for treatment.Reported event was confirmed by review of radiographs.Review of the available records identified the following: initial images demonstrate anatomic alignment of the reverse-type right shoulder arthroplasty complicated by glenosphere disassociation.Updated images demonstrate anatomic alignment.There was malalignment on the images with glenosphere disassociation as noted.No loosening, radiolucency or other abnormality.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-01751, 0001825034-2024-00218.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|