H3: the axium prime coil was returned for analysis.The axium prime coil was returned without its introducer sheath.Therefore, any contributing factors from the introducer sheath could not be assessed.The axium prime pushwire was found to be broken at break indicator and retained by release wire.In addition, the axium prime pushwire was found to be bent at ~142.5cm from proximal end.The axium prime implant coil was found to be detached from pushwire and was not returned for analysis.No other anomalies were observed.Based on the device analysis and the reported information, the customer¿s report of ¿coil resistance/stuck in sheath¿ could not be confirmed as the introducer sheath and implant coil were not returned.The root cause could not be determined.Regarding the damage to the axium prime pushwire, as the issue was not reported by the customer, it is possible the damage occurred during return to medtronic for evaluation.This regulatory report is being submitted as part of a retrospective review.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|