| Model Number |
021920 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Part 021920 natus embla sdx breakout - the doctor reported that the headbox fell on a patient's head because the rubber handle came off.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial report ref natus complaint# (b)(4).The patient was sleeping when the sdx box suddenly fell and hit the patient on the head.No injuries.Serial number of affected device not confirmed.Further investigation to be carried out.
|
| |
|
Event or Problem Description
|
|
Part 021920 natus brain monitor breakout - the doctor confirmed that the patient moved, pulled out the wires, and the breakout box fell on him.No injuries.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report 001 ref natus complaint#(b)(4).A natus specialist went onsite and checked each headbox and found that all removable handles are intact without rips or tears.Capa004628 was previously opened to investigate loose handles.Section d4 - serial number not provided.Per (b)(4).Xltek eeg psg risk analysis spreadsheet, hazard id - 6.13, severity 11 - critical, the risk is considered moderate.Per (b)(6).Complaint histories are reviewed routinely per quality system requirements and any complaint trends are assessed and documented as part of these reviews.Faillure confirmed: no investigation result code: neuro sbu|no issues noted closure rationale: complaint will be included in trending data for further review.
|
| |
|
Search Alerts/Recalls
|