| Catalog Number |
0240099155 |
| Medical Device Problem Codes |
No Display/Image (1183); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: loss of image.Probable root cause: overloading the cart, use error.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Search Alerts/Recalls
|