| Model Number |
SB-0520FC |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device was returned to olympus for evaluation.During inspection and testing, the allegations were confirmed.The tissue pad was worn and separated partially.The tissue pad was scorched.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
An olympus employee reported on behalf of a customer, during a therapeutic respiratory surgery when output several times the sonicbeat tissue pad got burnt.After resecting intercostal muscles twice, and trying to peel off the mucous member, it was found that the tip of pad had peeled off.The procedure was completed using a replacement device.There was no report of additional anesthesia, a procedural delay or patient harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Updated d4 to include additional supplemental lot number information.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a likely mechanism causing the reported event might be the following: the peeling of the tissue pad is likely to have occurred due to the fact that the tissue pad was worn and partly peeled off because the ultrasonic wave was output with the grasping part closed (including after the tissue was cut) without grasping the tissue.The following are the instructions for use which state: "do not activate output while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation".Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|