| Model Number |
VLSURGGEN |
| Medical Device Problem Codes |
Inaccurate Delivery (2339); Connection Problem (2900); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Superficial (First Degree) Burn (2685); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, an inspection was requested for possible burns along the edge part of the return electrode.Two weeks after the surgery, it became clear that the hair had fallen out and became like a keloid when the visit to the hospital was done for suture removal.The return electrode was being reused.Case was used on veterinary clinic.There was no patient involvement.From (b)(4): according to the reporter, 2 weeks after the surgery, the patient was brought to the hospital for suture removal and it was noted that the return electrode site became keloid-like (first degree burn).The keloid was not confirmed during and immediately after the procedure due to the animal's fur.No treatment was provided.It was noted during condition checking that the pediatric return electrode was reused and a gel was applied on the electrode part and the alarm of the return electrode contact status monitoring system did not sound.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h6, rfr/fdd adverse event to rfr/fdd rem alarm failure, rfr/fdd not detected and rem port issue, fdp/ime/imf first degree burn correction: e1(removed last name) d10 concomitant product: e7510-25, e7510-25 polyhesive ii ped ret el x25 (lot#unknown) medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, an inspection was requested for possible burns along the edge part of the return electrode.Two weeks after the surgery, it became clear that the hair had fallen out and became like a keloid (first degree burn) when the visit to the hospital was done for suture removal.It was noted that the return electrode site became keloid-like (first degree burn).The keloid was not confirmed during and immediately after the procedure due to the animal's fur.No treatment was provided.It was noted during condition checking that the pediatric return electrode was reused and a gel was applied on the electrode part and the alarm of the return electrode contact status monitoring system did not sound.Case was used on veterinary clinic.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device rem alarm failed, the device was not detected by the unit and the rem port on the generator had an issue.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|