Model Number 71005
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 01/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient was implanted with the nalu spinal cord stimulator system on (b)(6) 2021.After implantation the patient reported pain or discomfort at the midline incision.Investigation by the physician determined that the implanted anchors were too superficial.Surgical revision was performed on (b)(6) 2022 to move the anchors to a deeper location.Patient reports receiving therapy and no further discomfort.
 
Additional Manufacturer Narrative
No allegations of system or component failure.Patient discomfort is directly related to the location of the anchors and was corrected by moving the anchors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NALU NEUROSTIMULATION SYSTEM
Common Device Name
SPINAL CORD STIMULATOR
Manufacturer (Section D)
NALU MEDICAL INC
2320 faraday avenue
suite 100
calsbad CA 92008 2377
Manufacturer (Section G)
NALU MEDICAL, INC
2320 faraday ave
suite 100
carlsbad CA 92008 7241
Manufacturer Contact
terry villarba
2320 faraday avenue
suite 100
carlsbad, CA 92008-7241
7604482360
MDR Report Key17435179
Report Number3015425075-2023-00140
Device Sequence Number15656288
Product Code GZB
UDI-Device Identifier00812537033600
UDI-Public01008125370336001120120417231204
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K183579
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/04/2023
Device Model Number71005
Device Catalogue Number71005
Device Lot NumberA01631
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2022
Initial Report FDA Received Date08/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-