| Catalog Number |
NSLG2S35 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: (b)(6) 2023.B3: event year only reported: 2023.D4 batch #: x9647g.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a lash, very shortly after using the scissors, the surgeon noticed a squeaking noise.After pulling the device out of the trocar, the upper jaw broke off.No patient harm, no parts remain in patient.A new pair of scissors was opened and the surgery was successfully completed.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/26/2023.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/31/2023.D4 batch #: x96190.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was returned with the upper jaw detached and it was returned.In addition, the iblade was fractured and lifted.The device was connected to the generator and it was recognized.Because the jaw was detached not all functional testing could be performed with the generator.A probable cause of the damage could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information: it is possible that the noise heard occurred when the device was damaged.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch x96190, and no non-conformances were identified.
|
| |
|