MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507150MC |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Post Operative Wound Infection (2446)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Photo evaluation summary: upon visual evaluation of the image provided in the complaint no apparent damage or visual anomalies were observed.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the implant.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the implant is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.It should be noted that as part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, capa activity was not required.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: wound infection.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast augmentation with a 150cc mentor memorygel breast implant and experienced a wound infection on the left side post-operatively.As a result, the patient underwent explantation on an unspecified date.
|
| |
|
Additional Manufacturer Narrative
|
|
On august 18, 2023, the mentor failure analysis lab received the device for evaluation.A patient identifier was also confirmed.On august 24, 2023, mentor completed evaluation of the device.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the sm mod classic gel, 150cc returned device.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the implant.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the implant is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.It should be noted that as part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, capa activity was not required.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On september 19, 2023, mentor received a confirmed date of explant.Relevant fields have been updated.Manufacturer¿s reference number: (b)(4).
|
| |
|