KLS MARTIN SE & CO. KG LEVEL ONE CMF; SCREW Back to Search Results
Model Number 25-879-05-91
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 07/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A screw in a palatal expander footplate was found to be loose two months after implant and after successful consolidation had been achieved.It was removed along with the palatal expander.
 
Additional Manufacturer Narrative
An investigation was performed in the lab and there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.Review of the device history records was not possible due to no lot number being identified.The investigation results conclude that the root cause for this failure could not be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Corrections: d3 - manufacturer - establishment name only updated.G1 - all manufacturers - establishment name only updated.G1 - all manufacturers - manufacturing site for devices - establishment name and address - corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KLS MARTIN SE & CO. KG
kolbinger strasse 10
muehlheim/donau,
Manufacturer (Section G)
KLS MARTIN L.P.
p.o. box 16369
jacksonville FL 32245
Manufacturer Contact
melissa bachorski
kolbinger strasse 10
muehlheim/donau, de, FL 78570
GM   78570
9046417746
MDR Report Key17438236
Report Number9610905-2023-00036
Device Sequence Number6834826
Product Code HRS
UDI-Device Identifier00888118047755
UDI-Public(01)00888118047755
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K971297
Exemption Number2017029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-879-05-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer05/15/2024
Initial Report FDA Received Date08/01/2023
Supplement Report FDA Received Date06/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-