MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The complaint mr290v vented autofeed humidification chamber is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event or Problem Description
A healthcare facility in china reported that a mr290v vented autofeed humidification chamber was found leaking water before patient use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint mr290v vented autofeed humidification chamber was not returned to fisher & paykel healthcare for evaluation.Our investigation is therefore based on the information provided by the customer and our knowledge of the product.Results: the healthcare facility reported that a mr290v vented autofeed humidification chamber from f&p rt205 adult inspiratory heated breathing circuit was found leaking water before patient use.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
 
Event or Problem Description
A healthcare facility in china reported that a mr290v vented autofeed humidification chamber was found leaking water before patient use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Common Device Name
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17438782
Report Number9611451-2023-00714
Device Sequence Number2598234
Product Code BTT
UDI-Device Identifier09420012430410
UDI-Public(01)09420012430410(10)2102098741(11)220323
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K934140
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number2102098741
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2023
Supplement Date Received by Manufacturer08/07/2023
Initial Report FDA Received Date08/01/2023
Supplement Report FDA Received Date08/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
F&P RT205 ADULT BREATHING CIRCUIT.; F&P RT205 ADULT BREATHING CIRCUIT.