| Model Number |
BTB TIGHTROPE |
| Medical Device Problem Codes |
Failure to Advance (2524); Difficult to Advance (2920)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6)/2023, it was reported by an arthrex employee via email that a btb tightrope chinese mesh got stuck and would not advance.The surgeon could not unlock the mesh, and the button failed to rise.The case was completed using another tightrope.This was discovered during a procedure on (b)(6)2023.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.Upon evaluating the customer-provided photo, it was observed the suture loop system was broken.The sutures were frayed most likely due to excessive force.The most likely cause(s) of reported failure is user error following the surgical technique.According to dfu-0147.G.Precautions.1.Acl/pcl/btb/rt/abs tightrope only: excessive force on the shortening suture strands and/or tensioning the shortening suture strands not in-line with the bone socket may break the strands and impair ability to fully seat the implant.No additional force on the shortening suture strands is required when the graft/wedge construct reaches the desired position in the femoral socket and the graft stability is verified by pulling distally on the graft.
|
| |
|
Event or Problem Description
|
|
Additional information received on 1/2/2024: this was discovered during an acl procedure.
|
| |
|