• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BTB TIGHTROPE; SMOOTH FIXATION PIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BTB TIGHTROPE; SMOOTH FIXATION PIN Back to Search Results
Model Number BTB TIGHTROPE
Medical Device Problem Codes Failure to Advance (2524); Difficult to Advance (2920)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2023
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6)/2023, it was reported by an arthrex employee via email that a btb tightrope chinese mesh got stuck and would not advance.The surgeon could not unlock the mesh, and the button failed to rise.The case was completed using another tightrope.This was discovered during a procedure on (b)(6)2023.Additional information requested.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint allegation is confirmed.Upon evaluating the customer-provided photo, it was observed the suture loop system was broken.The sutures were frayed most likely due to excessive force.The most likely cause(s) of reported failure is user error following the surgical technique.According to dfu-0147.G.Precautions.1.Acl/pcl/btb/rt/abs tightrope only: excessive force on the shortening suture strands and/or tensioning the shortening suture strands not in-line with the bone socket may break the strands and impair ability to fully seat the implant.No additional force on the shortening suture strands is required when the graft/wedge construct reaches the desired position in the femoral socket and the graft stability is verified by pulling distally on the graft.
 
Event or Problem Description
Additional information received on 1/2/2024: this was discovered during an acl procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BTB TIGHTROPE
Common Device Name
SMOOTH FIXATION PIN
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key17441654
Report Number1220246-2023-07384
Device Sequence Number2735863
Product Code HTY
UDI-Device Identifier00888867019515
UDI-Public00888867019515
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K112990
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBTB TIGHTROPE
Device Catalogue NumberAR-1588BTB
Device Lot Number14543477
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2023
Supplement Date Received by Manufacturer01/02/2024
Initial Report FDA Received Date08/02/2023
Supplement Report FDA Received Date07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-