ZIMMER SURGICAL, INC. Z.S.G.M. CUTTER 2:1 RATIO ( CAUTION: SHARP ); SKIN GRAFT MESHER Back to Search Results
Model Number N/A
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the cutter failed the test cut.The cutter appeared to have even signs of wear throughout cutter consistent with intended use.Cutters cannot be fully repaired without replacement; therefore, it was returned to the customer as is.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.E1 telephone number: (b)(6) g2 country: united kingdom multiple mdr reports were filed for this event, please see associated reports: 0001526350-2023-00322-1, 0001526350-2023-00834, 0001526350-2023-00838, 0001526350-2023-00840.
 
Event or Problem Description
It was reported that during a demonstration, the device seemed too tough to use and was tearing the donor skin.There was nothing noticeably broken on the mesher and it seemed to be assembled correctly but it was tearing the skin as it was put through the mesher on the carrier.During product evaluation, the cutter failed the test cut.There was no patient involvement.Due diligence is complete and there is no additional information available.There was no adverse event associated with this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Z.S.G.M. CUTTER 2:1 RATIO ( CAUTION: SHARP )
Common Device Name
SKIN GRAFT MESHER
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key17442738
Report Number0001526350-2023-00836
Device Sequence Number2905614
Product Code FZW
UDI-Device Identifier00889024375451
UDI-Public(01)00889024375451(11)190426(10)64371102
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
PREAMENDEMEN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00770302000
Device Lot Number64371102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2023
Initial Report FDA Received Date08/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-