Investigation: visual inspection: during the visual inspection a deformation of the outer housing of the prosa valve could be determined.The measurement of the plane parallelism could confirm that with a value of -0,03 mm - outside tolerance (0 ± 0.02 mm).Permeability test: a permeability test has indicated that the prosa is permeable.During the permeability test some particles.Computer controlled test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in the vertical positions.The results show that the valve is operating within the accepted tolerance in the vertical position.Adjustment test: the prosa valve was tested and is not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Internal inspection : after dismantling of the valve, deposits were found in prosa.Results: based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|