MAUDE Adverse Event Report: MEDTECH SAS ROSA ONE BRAIN; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SAS ROSA ONE BRAIN; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number N/A
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The subject device was evaluated by a fse on-site.It was reported that the fse upgraded the software and replaced the cpu board of the controller.The system was operational, and all functional checks passed after this visit.As the issue was determined to be related to hardware, a review of the log files was not performed.Review of the device history record found the below which could be related to the reported event: there was a previous corrective maintenance on the same device during which the fse found the cpu board of the controller needed to be replaced.This issue is referenced in (b)(4), the cpu board on the rose one system was malfunctioning so a loaner unit was used in the field until investigation/repair could be completed.(b)(4) is the corrective maintenance on this cpu.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported during corrective maintenance, controller will not connect to computer.No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE BRAIN
Common Device Name
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SAS
432 rue du rajol
mauguio, languedoc-roussillon 34130
FR  34130
Manufacturer (Section G)
MEDTECH SAS
432 rue du rajol
mauguio, languedoc-roussillon 34130
FR   34130
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key17445794
Report Number3009185973-2023-00019
Device Sequence Number2531725
Product Code HAW
UDI-Device Identifier03760244032492
UDI-Public(01)03760244032492(11)180605
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K200511
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberROSAS00203
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2023
Initial Report FDA Received Date08/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-