| Model Number |
Ø7X 30MM BC IF SCRW, VENTED |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the screw fractured while being inserted.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update avoe (b)(6) 2023.It was confirmed that no broken parts remained inside the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint was confirmed.One unpackaged ar-4030c-07 serial/batch (b)(6) was returned for investigation.The returned interference bio screw was visually inspected and noted that it was broken.Damage was observed inside the screw.The inside geometry was lost.The method of bone preparation employed during the procedure, as well as the bone quality encountered, was not provided.The most likely cause is continually applying torque when the implant is seated and /or not advancing.Per dfu-0111 at revision 1.G.Precautions.6.Bio-absorbable interference screw only: it is important to completely seat the screwdriver to prevent potential stripping of the hex and/or screw fracture during insertion or removal.
|
| |
|
Search Alerts/Recalls
|