| Model Number |
880-307/11 |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Screws that attach augment to fem component were put into augments backwards.Screws should be fed from prox to distal.There were fed through the augment from distal to proximal thus causing the surgeon to have to remove them and reinsert them.The patient in no way was affected and the surgery proceeded as planned.Dr.S.Does not want or need any follow up.Has specifically asked to just inform link so they can correct the problem moving forward.Customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Recall class ii as per information of (b)(6) 2023 and therewith classified as reportable.
|
| |
|