| Catalog Number |
129435315 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Joint Laxity (4526)
|
| Date of Event |
07/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
This patient had revision sigma components currently implanted but the tibial component had fallen into varus and the tibial stem was protruding out of the lateral cortex of the tibia.This patient also had stability issues so surgeon removed both the tibial and femoral components and replaced them with an attune revision tibial tray, stem, and sleeve.Doi: unknown.Dor: (b)(6) 2023.Affected side: right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: b5, d10 concomitant.
|
| |
|
Event or Problem Description
|
|
Additional information received states that there was nor surgical delay.
|
| |
|