| Date of Event |
07/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during procedure.All pieces were retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: 90 s max cruise that snapped inside the patient the failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be excessive force used when inserting or removing the probe, a probe inserted into an obstructed passageway, or any forceful impact to the tip of the probe with rigid objects.The probe is used as a tool for a mechanical displacement of tissue or bone, or to pry or scrub.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|