• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. NV GXL LNR, 10 DEG FACE, 36MM ID, GRP 2 CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. NV GXL LNR, 10 DEG FACE, 36MM ID, GRP 2 CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER Back to Search Results
Catalog Number 138-36-52
Medical Device Problem Codes Naturally Worn (2988); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 01/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.
 
Event or Problem Description
As reported via legal documentation, a patient had left hip replacement surgery on (b)(6) 2009.They underwent left hip revision surgery on (b)(6) 2021, approximately 11 years 1 month post primary procedure.There is no other patient demographic or medical history available.There is no information on the surgical procedure or patient outcome.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
Concomitants: 1512114 120-65-15 - bone screw 6.5mm dia x 15mm long 1578827 120-65-20 - bone screw 6.5mm dia x 20mm long 1562039 142-36-00 - cocr fem head 36mm +0 offset 12/14 1415514 160-30-11 - pf spline plasma w/ha sz 11 1361874 180-01-52 - nv crown cup clstr hole 52mm group 2 investigation results - based on the available information, the patient involved meets the following risk criteria specified in the hhe: implanted with a lateralized liner and combination of the largest available femora head and/or thinnest available acetabular liner was used.The most likely underlying cause for the revision reported is a combination of risk factors specified in hhe-2020-09-11-02 and being implanted for more than 10 years.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LNR, 10 DEG FACE, 36MM ID, GRP 2 CUP
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
miguel sosa
3523782617
MDR Report Key17446733
Report Number1038671-2023-01849
Device Sequence Number17987736
Product Code LZO
UDI-Device Identifier10885862024886
UDI-Public10885862024886
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K070479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date11/09/2013
Device Catalogue Number138-36-52
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer12/12/2023
Initial Report FDA Received Date08/02/2023
Supplement Report FDA Received Date12/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1729-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-