| Catalog Number |
0502555400 |
| Medical Device Problem Codes |
Material Separation (1562); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that screw came off during procedure.The screw was successfully removed.
|
| |
|
Event or Problem Description
|
|
It was reported that screw came off during procedure.The screw was successfully removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib, lisa, biomed,broken, po, sent ack letter.Update: during the procedure we were using the cup when they stopped working and a screw came into the field of view.The screw came off inside the patient¿s uterus during the procedure.The procedure was completed and the screw was retrieved.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be rough handling and/or contact with metal objects during processing and sterilization, causing components to become loose or broken.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Search Alerts/Recalls
|