This report has been identified as b.Braun medical internal report number (b)(4).One used catheter with packaging and one unused catheter without packaging were returned for evaluation.Visual evaluation of the samples did show the tip of the used catheter to be sheared/broken off.The second catheter contained the tip and was intact.Although the used catheter was sheared, since it was not broken or damaged upon the opening of the kit and it had been used in a procedure, it is not confirmed to be the result of any manufacturing process.Per the manufacturer's investigation this defect is not likely to have happened during the manufacturing process.It is believed to happen during application.The user will be referred to the ifu which states, precaution: advancing the catheter more than 5 cm past the needle tip may increase the likelihood of kinking or knotting.Caution: do not withdraw the catheter through the epidural needle due to the possibility of shearing or kinking the catheter.When removing the catheter, excessive force should never be applied.If the catheter becomes difficult to withdraw, consult standard textbooks for specific techniques." review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|