• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION HABIB ENDOHPB; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION HABIB ENDOHPB; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00500070
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: impact code f1001 is being used to capture the reportable issue of aborted/cancelled procedure.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2023 that a habib endohpb rf catheter was used to treat an inoperable ampullary malignant tumor during a radiofrequency ablation procedure performed on (b)(6) 2023.The lesion was not dilated prior to the first attempt at ablation.During the procedure, the device was unable to deliver energy and the lesion could not be ablated.Furthermore, there was no blanching of the tissue noted.The procedure was not completed due to the availability of the device.There were no patient complications reported as a result of this event.The patient's condition following the procedure was reported to be stable.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2023 that a habib endohpb rf catheter was used to treat an inoperable ampullary malignant tumor during a radiofrequency ablation procedure performed on (b)(6) 2023.The lesion was not dilated prior to the first attempt at ablation.During the procedure, the device was unable to deliver energy and the lesion could not be ablated.Furthermore, there was no blanching of the tissue noted.The procedure was not completed due to the availability of the device.There were no patient complications reported as a result of this event.The patient's condition following the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: impact code f1001 is being used to capture the reportable issue of aborted/cancelled procedure.Block h10: a habib endohpb rf catheter was received for analysis.Analysis of the returned device found the tube strongly damaged.Both electrodes were noted to have worked (glue has turned to yellow color); however, there is no connection noted at the proximal electrode.No other problems with the device were noted.Product analysis confirmed the reported event of device failure to deliver energy.It is most likely that the reported event was due to the observed event of catheter bent/kinked and that the device was damaged during use.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to procedure.A product labeling review identified that the device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HABIB ENDOHPB
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
STATICE MANUFACTURING
9 rue thomas edison
besancon 25000
FR   25000
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17449970
Report Number3005099803-2023-04079
Device Sequence Number12604218
Product Code KNS
UDI-Device Identifier08714729981909
UDI-Public08714729981909
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K231533
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/19/2025
Device Model NumberM00500070
Device Catalogue Number0007
Device Lot Number0578810010
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2023
Supplement Date Received by Manufacturer09/27/2023
Initial Report FDA Received Date08/03/2023
Supplement Report FDA Received Date10/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
Patient Weight65 KG
-
-