Model Number 131F7
Health Effect - Clinical Code Pain (1994)
Date of Event 07/21/2023
Type of Reportable Event Malfunction
Event or Problem Description
Elderly male with past history of cardiac disease.Procedure: right heart cath and stenting.7f swan-ganz was inserted via the 7f sheath in the right femoral vein.There was difficulty advancing the swan.The x-ray was panned down to the right groin area and a knot in the swan.The decision was made to remove the sheath and swan together in attempt to untie the knot.The swan broke into two pieces as the sheath was removed.Vascular surgeon performed cutdown and removed the knotted swan under local anesthetic.Patient discharged the next day without known harm.Difficulty obtaining right femoral venous access.Once access was obtained, a 7f sheath was inserted.The 7f swan-ganz was inserted via the 7f sheath in the right femoral vein.There was difficulty advancing the swan.An 0.25 wire was inserted to help advance the swan.The swan was not advancing.The x-ray was panned down to the right groin area and a knot in the swan was seen.An attempt was made to untie the knot.The decision was made to remove the sheath and swan together in attempt to untie the knot.The swan broke into two pieces as the sheath was removed.A portion of the swan was sticking out of the patient's skin.The knot was still in the right femoral vein.Access was obtained in the left femoral vein and a 10f sheath was inserted.A snare was inserted via the 10f sheath in the left femoral vein in an attempt to remove the knot.The snare was able to grab the knot, but the knot was unable to move across to the left groin.This was too painful for the patient.A 7f sheath was inserted in the right femoral vein over the piece of swan that was sticking out of the patient's skin.Dr 1, vascular surgeon, entered the procedure.He exchanged the 7f sheath in the right femoral vein for a 16f sheath.The snare was re-inserted in the 16f sheath in the right femoral vein.The knot could not be removed via the 16f sheath.The knot could not be removed via the 16f sheath.The cvor (cardiovascular operating room) team entered the procedure.Dr 1 was needed in a case in the hybrid or.Dr2, vascular surgeon, entered the case to relieve dr 1.Dr 2 performed a vascular cut down on the right femoral vein and the swan knot was retrieved.This was all done under local anesthetic to the right groin.Manufacturer response for catheter, intravascular, diagnostic, swan-ganz (per site reporter), will obtain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
one edwards way
irvine CA 92614
MDR Report Key17449980
Report Number17449980
Device Sequence Number18015303
Product Code DYG
UDI-Device Identifier00690103000405
UDI-Public(01)00690103000405(17)250509(11)230510(10)65067168
Combination Product (Y/N)N
Initial Reporter StateOH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 07/21/2023
Report Date (Section F) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number131F7
Device Catalogue Number131F7P
Device Lot Number65067168
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer08/03/2023
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/03/2023
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
Date Report Sent to FDA07/28/2023
-
-