• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND MODULAR HUM STEM D12 HA; DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND MODULAR HUM STEM D12 HA; DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS Back to Search Results
Catalog Number 130712000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Date of Event 07/20/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Patient was revised due to loosening of metaglene and glenosphere.Screws were also broken.Patient also experienced pain and possible infection.There was no surgical delay.Doi- (b)(6) 2012.Dor- (b)(6) 2023.Affected side- right shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : patient was revised due to loosening of metaglene and glenosphere.Screws were also broken.Patient also experienced pain and possible infection.There was no surgical delay.Doi- (b)(6) 2012, dor- (b)(6) 2023, affected side- right shoulder.The photo was returned to depuy synthes for evaluation.The attached photo does not depict the reported dxtend modular hum stem d12 ha.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.The overall complaint was not confirmed as the photograph provided is not representative of the reported complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND MODULAR HUM STEM D12 HA
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17450257
Report Number1818910-2023-15861
Device Sequence Number7347808
Product Code KWS
UDI-Device Identifier10603295027171
UDI-Public10603295027171
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number130712000
Device Lot Number5126249
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2023
Supplement Date Received by Manufacturer09/14/2023
Initial Report FDA Received Date08/03/2023
Supplement Report FDA Received Date09/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/18/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND GLENOSPHERE STD D38MM; DXTEND METAGLENE; DXTEND MOD EPI 1 ECC LEFT HA; DXTEND MODULAR HUM STEM D12 HA; DXTEND SCREW LOCK D4.5X24MM; DXTEND SCREW LOCK D4.5X24MM; DXTEND SCREW NO LOCK D4.5X24MM; DXTEND SCREW NO LOCK D4.5X42MM; DXTEND STAND PE CUP D38 +6MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight102 KG
-
-