|
A review of the device history records for the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation, "excessive force used" , "device used at sub optimal indications", "device forced against resistance", "material fatigue", "excessive force", "operational context", "inexperienced user", "unanticipated user error" and/or "improper handling" appears to have impacted on the event as reported.
|
|
Nature of complaint: after the procedure with the phoenix catheter, he removed the phoenix without complications.However, when he attempted to remove the wire from the patient, he experienced a slight resistance for a short period of time, and then he saw outside the patient that the tip of the wire was missing.The tip of the wire was left in a side branch of the artery in the lower leg.
|
|
As reported; "after the procedure with the phoenix catheter, he removed the phoenix without complications.However, when he attempted to remove the wire from the patient, he experienced a slight resistance for a short period of time, and then he saw outside the patient that the tip of the wire was missing.The tip of the wire was left in a side branch of the artery in the lower leg." the sample was returned, placed in a double zip-lock type bag marked biohazard.The sample was then placed into 2 boxes which were protected with foam.The guidewire returned coiled in a circular shape.A visual and tactile examination of the returned guidewire was completed, and the following features were observed.A visual inspection of the returned guidewire confirmed that the guidewire was fractured.(core & coil).The distal tip was not returned.The proximal joint was intact, and the coil was stretched.This guidewire is a 2- piece construction: coil and core.A microscopic examination of the distal section of the returned guidewire revealed a sharp point the core, where it was sheared.There were body bends noted on the returned guidewire, close to the distal section.The body of the core had a hard, tissue like substance attached when returned.This substance was located approximately 45cm from the distal tip.No other damage was noted on the returned product.The sample was sent for external analysis.The sme analysis of this fracture reveals the following primary features.This is a bending overload (or combined bending-tensile overload) fracture.You can see the sharp bend immediately adjacent to the fracture.The dimpled fracture surface indicates a ductile (rather than brittle) fracture.I did not see any foreign materials or pre-existing mechanical damage that would have contributed to the fracture.There were some small "smudges" of pd-re from where the edge of the fracture and the od surface rubbed against the id of the radiopaque coil, but these would not have contributed to the fracture in any way.The eds confirms that the wire material is nitinol.Overall, it looks like the wire got bent to the point of kinking during use and this led to fracture at this point.A review of the device history records for the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided by the supporting documentation, "excessive force used" , "device used at sub optimal indications", "device forced against resistance", "material fatigue", "excessive force", "operational context", "inexperienced user", "unanticipated user error" and/or "improper handling" appears to have impacted on the event as reported.The sample was returned, placed in a double zip-lock type bag marked biohazard.The sample was then placed into 2 boxes which were protected with foam.The guidewire returned coiled in a circular shape.A visual and tactile examination of the returned guidewire was completed, and the following features were observed: · a visual inspection of the returned guidewire confirmed that the guidewire was fractured.(core & coil).The distal tip was not returned.The proximal joint was intact, and the coil was stretched.· this guidewire is a 2- piece construction: coil and core.· a microscopic examination of the distal section of the returned guidewire revealed a sharp point the core, where it was sheared.· there were body bends noted on the returned guidewire, close to the distal section.· the body of the core had a hard, tissue like substance attached when returned.This substance was located approximately 45cm from the distal tip.· no other damage was noted on the returned product.A dimensional check was performed to ensure critical guidewire dimensions are within specification per vol#003g guidewire drawing.The following dimensions were observed: · outer diameter guidewire - pass - 0.0129" (specification: min: 0.0126'' - max:0.0134") · overall length of the guidewire could not be measured as the distal tip was not retuned.The length of the core was measured as seen below in core dimensional check.The required length of the guidewire is 300cm +/-2cm.· overall diameter of the proximal joint - pass - 0.0132" (specification: max: 0.0142") a dimensional check was performed to ensure critical core dimensions are within specification per vol#003r core drawing.The following dimensions were observed: · outer diameter core body - pass - 0.0129" (specification: min: 0.0126'' - max:0.0134") · outer diameter core tip, tapered section - 0.00690" · length of core from the proximal to the fracture point - 296.6cm.(specification: 300 +/- 2cm).· core length from the proximal joint to the fracture point measured approximately 1.8cm.According to the guidewire drawing vol#003g the length of the mandrel from proximal joint to the distal tip should be 5.5 cm, which reflected approximately 3.7cm of distal section being removed.· the core diameter at fracture point measured 0.00690''.Which confirms the fracture is on the tapered core section as per vol#003r drawing.The fracture occurred on the core tapered section approximately 3.7cm from the distal tip.Only the proximal end of the broken wire was returned.The fracture was characterized by sme at the integer salem met lab under metallographic test report.The sme analysis of this fracture reveals the following primary features: this is a bending overload (or combined bending-tensile overload) fracture.You can see the sharp bend immediately adjacent to the fracture.The dimpled fracture surface indicates a ductile (rather than brittle) fracture.I did not see any foreign materials or pre-existing mechanical damage that would have contributed to the fracture.There were some small "smudges" of pd-re from where the edge of the fracture and the od surface rubbed against the id of the radiopaque coil, but these would not have contributed to the fracture in any way.The eds confirms that the wire material is nitinol.Overall, it looks like the wire got bent to the point of kinking during use and this led to fracture at this point.
|
|
Nature of complaint: after the procedure with the phoenix catheter, he removed the phoenix without complications.However, when he attempted to remove the wire from the patient, he experienced a slight resistance for a short period of time, and then he saw outside the patient that the tip of the wire was missing.The tip of the wire was left in a side branch of the artery in the lower leg.
|