CONMED UTICA ELEC,BLADE,4",SMOKE EVAC,40; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
138112 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During incoming inspection, the distributor rejected this device, 138112, for an insufficient heatseal.The device was encroaching the seal.There was no contact with the patient as this was found during incoming inspection.This report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Received nine 138112 in original packaging.Lot number was verified.Performed a visual inspection, the device is encroaching the seal.Performed a functional inspection, the devices were dye leak tested which indicated that the packaging did have an insufficient heat seal.Root cause cannot be determined, however, based upon evaluation of the device, a possible cause of this event is the device was encroaching into the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised to inspect and test each device before each use.We will continue to monitor for trends through the complaint system to assure patient safety.
|
| |
|