|
The reported issue was confirmed.The root cause of the reported issue is that the optical sensor is not situated correctly in the housing causing it to read incorrectly.The unit was evaluated and the reported issue was confirmed as it was noted that the gasket foam needs to be added to optical sensor of the sensica device.The manufacturing oi sccs1002x was updated.Applied gasket foam to optical sensor.Completed final functional testing.A dhr is not required as the device has undergone previous servicing and therefore the reported event is not manufacturing related.The instructions for use were found adequate and state the following: "indications for use: indications: the bd sensica¿ urine output system is an automated system for continuous monitoring of urine output (uo) and core bladder temperature when connected to a temperature-sensing foley catheter.Target population: the bd sensica¿ urine output system is intended for bedside monitoring of urine output for any patient with an indwelling, urological catheter, drainage tubing and collection bag, typically in critical care settings or where close monitoring of urine output is desired.The bd sensica¿ urine output system is also intended for monitoring core bladder temperature when used with the bd sensica¿ temperature monitor module and a bd 400 series temperature-sensing foley catheter or equivalent.Contraindications: there are no known contraindications for use with patients who have indwelling, urological catheters in place.General warnings: 1.Medical electrical equipment requires special precautions regarding electromagnetic compatibility (emc) and needs to be installed and put into service according to the emc information provided in the charts at the end of these instructions for use.2.Portable and mobile radiofrequency (rf) communications equipment can affect medical electrical equipment.3.The bd sensica¿ urine output system should not be stacked with other equipment.4.The bd sensica¿ urine output system has a degree of protection against electric shock of applied parts classified as type bf.This device is classified as an iec class i device.5.When using class 1 xp power model number: alm65us12 power supply, to avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth.6.Do not immerse or submerge the bd sensica¿ urine output stand, display monitor, ring, temperature monitor module, temperature sensor cables, or monoplug adapter in water or other liquids.Do not pour liquids over the display monitor, ring, temperature monitor, cables or monoplug adapter.If liquids accidentally spill onto the device(s), wipe off liquid with soft cloth as soon as possible.See section 14 for complete care instructions.7.This device is not suitable for use in the presence of flammable mixtures.8.This device is not suitable for use in oxygen rich environments.9.The use of transducers and cables other than those specified, with the exception of transducers and cables sold as replacement parts, may result in increased emissions or decreased immunity of the bd sensica¿ urine output system.10.The bd sensica¿ urine output system is designed to be used with any bard¿ foley catheter and urological disposables connected to any standard urine drainage bag (2000 or 2500 ml), with or without a urine meter.11.If using a 400-series temperature-sensing foley catheter, see product instructions for use for mri compatibility.Refer to foley catheter instructions for use for indications and further product information.12.If using a 400-series temperature-sensing foley catheter, the operator is responsible for ensuring the compatibility of the temperature sensing foley catheter to the bd sensica¿ urine output system and its accessories.Incompatible components can result in degraded performance.13.No modification of the bd sensica¿ urine output system or any of its accessories is allowed.Warning: this product can expose you to di(2-ethylhexyl) phthalate (dehp), which is known to the state of california to cause cancer and birth defects or other reproductive harm.For more information, go to: https://www.P65warnings.Ca.Gov.Operation warnings: 1.Federal law restricts this device to sale or use by or on the order of a licensed physician.2.The bd sensica¿ urine output system should only be used on the request by and under the direction of a licensed physician.3.Do not use bd sensica¿ ring (disposable) if product appears broken.4.The ¿accuracy check¿ should be performed prior to first use, when there is a question about system accuracy, and/or after any servicing or repairs to ensure accurate measurements.Always use a new bd sensica¿ ring to complete an accuracy check.If the accuracy check fails call bd customer service at (b)(6).This self-test takes about one minute and can be accessed from the new patient and settings screen or from the administrator settings.5.Do not change time and date settings during system operation.Time and date settings should be set when system is initially installed or between monitoring sessions, if needed.6.Ensure that the urine drainage bag hanging on the bd sensica¿ ring is always below the level of the patient¿s bladder to maintain continuous, gravity flow of urine into the collection bag.When connecting the single patient use bd sensica¿ ring to the system stand, use a firm, clockwise twisting motion.In order to avoid damaging components, do not apply excessive force or torque to the ring or the system¿s ring interface when connecting the device.7.To properly stabilize the system for accurate urine output measurement, always use the bd sensica¿ tube band positioning strip provided in the bd sensica¿ ring kit to correctly position the tube band on the drainage tubing.Insert the drainage tubing in the tube holder with the tube band in the tube holder window.The band should be permanently attached to the drainage tubing using the provided peel-off adhesive backing.When removing the bag from the ring for any reason, simply pull the tubing out of the tube holder.When re-hanging the bag, reinsert the tubing into the tube holder with the tube band in the tube holder window.Do not twist or kink tubing at any time during use.The tube band positioning strip does not need to be applied when using a drainage bag that comes with pre-marked tubing.8.Variations in the connection of the drainage tube to the tube holder or the bag to the ring can impact system accuracy.It is advised to not disturb the components of the bd sensica¿ urine output system while in use unless required for patient care.9.During operation, keep device stand level on the floor to avoid tipping the unit over and damaging system components.Additionally, the device stand should be kept level during a monitoring session as tilting of the device may impact the accuracy of urine output measurements.Use the crossbar handle on the back of the device to move the bd sensica¿ urine output system.10.Do not pick up the device from the front of the system.Use caution not to drop the system while moving the system.11.Take care not to trip over the bd sensica¿ urine output system or the power cord to avoid user injury or accidentally pulling out the indwelling catheter.12.Do not set alarm limits to extreme values that can render the alarms useless.Use only the bd sensica¿ ring when using the bd sensica¿ urine output system for correct operation.13.Do not lean on, pull on, or allow anything to rest on the urine collection bag, bd sensica¿ ring or drainage tubing while the urine bag is hanging on the bd sensica¿ urine output system.This will generate a system alarm, could cause erroneous urine output readings, and may break the load cell measurement assembly.Breakage due to excessive force/weight applied to the ring or load cell assembly will constitute misuse of the system and repairs will not be covered under warranty.14.Do not sterilize any component of the bd sensica¿ urine output system.To clean the display touch screen, wipe with a soft cloth or use a germicide wipe such as super sani-cloth®.Do not spray harsh cleaners directly on the display touch screen.See section 14 for complete care and maintenance of the bd sensica¿ urine output system.15.The bd sensica¿ ring is a single patient use device.After ending patient monitoring, disinfect ring using facility germicide wipes.After use, the ring may be a potential biohazard.Handle and dispose in accordance with accepted medical practice, facility procedures, and applicable local, state, and federal laws and regulations.16.Do not attempt to replace the system¿s battery.Contact bd customer service at (b)(6) to arrange for product service.17.Contact bd customer service at +1.844.823.5433 regarding serviceable components in the bd sensica¿ urine output system.Do not attempt to open control covers or repair the bd sensica¿ urine output system without proper training and approved replacement kits.To do so, may alter urine output measurements and void the warranty.Contact customer service to report system problems and arrange for product service.Note: there are no serviceable parts in the temperature monitor module.Do not attempt to open the bd sensica¿ temperature monitor module housing.With the bd sensica¿ temperature monitor module, bd sensica¿ temperature sensor cable and monoplug adapter, only use a bd 400 series temperature-sensing foley catheter or equivalent.18.Do not attempt to use the bd sensica¿ urine output system with any criticore¿ disposable products.After use, the bd sensica¿ urine output system, temperature monitor module, temperature sensor cable, and monoplug adapter may be a potential biohazard.Clean and disinfect the products before disposal.Handle and dispose of the products in accordance with accepted medical practice, facility procedures, and applicable local, state, and federal laws and regulations.These products contain electrical and/or electrical equipment, and the bd sensica¿ urine output system (sccs1002 and sccs1002e) contains a lithium-ion battery that must be disposed of per local laws and regulations.19.Do not simultaneously touch the patient and the visible metal contacts of the temperature monitor module or the ring interface.Precautions: 1.During system start up and in general practice, plug the bd sensica¿ urine output system into a wall power supply whenever possible.The system screen will dim when the system is unplugged to maximize battery life.After using the system on battery back-up, plug it back into the wall power supply for recharging and to avoid system shut down due to a drained battery.2.After removing a patient from urine output monitoring with the bd sensica¿ urine output system, shut down the system completely to avoid battery drainage while not in use.3.Low priority alarms may interrupt urine output monitoring and require user action to resume normal urine output monitoring functions.4.If the power cord icon is displayed on the screen, the power supply is not connected.5.If the system is repositioned, always ensure that the power supply is fully seated on both ends.6.Do not disturb the system during stabilization." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was evaluated.
|