• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSTITUT STRAUMANN AG S, Ø 3.3 MM, RN, SLA 14.0 MM; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSTITUT STRAUMANN AG S, Ø 3.3 MM, RN, SLA 14.0 MM; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 043.134S
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Fibrosis (3167)
Date of Event 06/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted on (b)(6) 2021 in fdi 36.On (b)(6) 2022, non-osseointegration was verified.The device was forwarded to the manufacturer.At the event the patient experienced: infection, mobility and inflammation.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the non-integration of an endosseous dental implant during the healing phase is a known inherent risk of the treatment with dental implants.Most implant failures occur before occlusal loading (analainen et al.2009).Based on clinical studies about 1-3% of the implants fail within the first year after implantation (ganeles et al.2008).Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported early failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S, Ø 3.3 MM, RN, SLA 14.0 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key17453216
Report Number0009613348-2023-084588
Device Sequence Number15299372
Product Code DZE
UDI-Device Identifier07630031718617
UDI-Public07630031718617
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K983742
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/16/2022
Device Catalogue Number043.134S
Device Lot NumberMT307
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/03/2023
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-