• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MO016 MO-MOBERLY INFANT HEEL WARMER SQUEEZE 4X4IN; INFANT HEEL WARMER (CHEMICAL HEAT PACK)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MO016 MO-MOBERLY INFANT HEEL WARMER SQUEEZE 4X4IN; INFANT HEEL WARMER (CHEMICAL HEAT PACK) Back to Search Results
Model Number 11460-010T
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
It has been reported that a laboratory support technician was activating the heel warmer by squeezing it, the top of the heel warmer busted open.There was no injury/adverse event.The pack material got on the clinician's clothes.There was no skin contact.It was washed away with soap and water.The clinician was not harmed.
 
Additional Manufacturer Narrative
Two samples were received for evaluation and were visually inspected for any defects.The samples were already activated and a tear was noted in pouch.The root cause which contributed to the tear was due to a machine misalignment.The device history record of the complaint lot was reviewed.The lot was inspected and released in compliance with all requirements.No anomalies were found during review of the records.Cardinal health will continue to monitor and trend all similar reported product related issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT HEEL WARMER SQUEEZE 4X4IN
Common Device Name
INFANT HEEL WARMER (CHEMICAL HEAT PACK)
Manufacturer (Section D)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer (Section G)
MO016 MO-MOBERLY
808 w highway 24
moberly MO 65270
Manufacturer Contact
patricia tucker
3651 birchwood dr.
waukegan, IL 60085
8478874151
MDR Report Key17454208
Report Number1423537-2023-00974
Device Sequence Number12580112
Product Code MPO
UDI-Device Identifier10630140017391
UDI-Public10630140017391
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11460-010T
Device Catalogue Number11460-010T
Device Lot NumberV3D237
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Initial Report FDA Received Date08/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
-
-