| Catalog Number |
06407137190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter questioned a high result for 1 patient tested for crej2 creatinine jaffé gen.2 (crej2) on a cobas 8000 c 702 module.The initial result was 4.01 mg/dl.This result was reported outside of the laboratory where the physician requested a re-check.A 2nd sample was obtained and the result was 0.28 mg/dl.This sample was repeated with a 1/5 dilution and the result was 0.28 mg/dl.On 11-jul-2023 the original sample was repeated with a result of [-] 0.32 mg/dl.This sample was repeated with a 1/5 dilution and the result was 0.35 mg/dl.On 12-jul-2023 a 3rd sample was obtained and the result was 0.16 mg/dl.The low results were believed to be correct based on the clinical status of the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Section e1 facility name continued: reception centr quai de dechargement 2 the investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The c702 module serial number was (b)(6).The customer used 13 mm primary sample tubes with no rack adapters.Calibration and qc were acceptable.An abnormal aspiration (sample probe) error was observed on the alarm trace data from the day of the event.The customer did not provide any additional information and sample material is no longer available.A general reagent and hardware issue were excluded as only one crej2 result was in question and no other results were affected.Based on the data available, the investigation determined the event was consistent with a pre-analytical issue.The investigation did not identify a product problem.The specific cause of the event could not be determined.
|
| |
|
Search Alerts/Recalls
|