| Model Number |
MON-000005-01 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a recurrent fault state occurred during initial system registration of the monarch bronchoscopy system, preventing the use of the system.The diagnostic procedure was aborted.The patient had already been anesthetized.No clinical consequences to the patient were reported due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A system log review confirmed the reported event.The cause of the recurrent system fault was isolated to the arms of the monarch bronchoscopy system.The monarch bronchoscopy system was repaired by replacing the affected arms.A device history record (dhr) review for the system was performed and no non-conformances were found that relate to the reported issue.No further action is required at this moment as issue is a known failure mode.This failure will continue to be tracked in regular trend review meetings.
|
| |
|
Search Alerts/Recalls
|