Model Number MON-000005-01
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a recurrent fault state occurred during initial system registration of the monarch bronchoscopy system, preventing the use of the system.The diagnostic procedure was aborted.The patient had already been anesthetized.No clinical consequences to the patient were reported due to this event.
 
Additional Manufacturer Narrative
A system log review confirmed the reported event.The cause of the recurrent system fault was isolated to the arms of the monarch bronchoscopy system.The monarch bronchoscopy system was repaired by replacing the affected arms.A device history record (dhr) review for the system was performed and no non-conformances were found that relate to the reported issue.No further action is required at this moment as issue is a known failure mode.This failure will continue to be tracked in regular trend review meetings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH PLATFORM
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID) AND ACCESSORIES
Manufacturer (Section D)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA 94065
Manufacturer (Section G)
AURIS HEALTH, INC.
150 shoreline drive
redwood city CA
Manufacturer Contact
monica nachbar
150 shoreline drive
redwood city, CA 94065
6502085208
MDR Report Key17455864
Report Number3014447948-2023-00014
Device Sequence Number16821686
Product Code EOQ
UDI-Device IdentifierB634MON000005010
UDI-Public+B634MON000005010/16D20210706
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K211493
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMON-000005-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer07/06/2023
Initial Report FDA Received Date08/03/2023
Supplement Report FDA Received Date09/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-