• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC BD BD® WHITACRE SPINAL NEEDLE AND BD¿ WEISS EPIDURAL NEEDLE TRAY; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC BD BD® WHITACRE SPINAL NEEDLE AND BD¿ WEISS EPIDURAL NEEDLE TRAY; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 405828
Medical Device Problem Codes Crack (1135); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D.4 device expiration date: unknown.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.H.4.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that the bd bd® whitacre spinal needle and bd¿ weiss epidural needle tray was cracked causing leakage.The following information was provided by the initial reporter: we are having major issues with the connectors for the catheter in our current cse kit.They crack and epidural solution is delivered to the bed instead of the epidural space.Please advise on next steps to correct the issue as this is a patient safety concern.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 02-oct-2023 h6: investigation summary eight samples were received by our quality team for investigation.Upon visual inspection, it was observed that the cracks were present at the luer end of the connector therefore, the reported failure mode was confirmed.A device history review could not be completed as no batch number was provided.Based on the available information, the root cause of the issue corresponds to a supplier defect, a quality notification to the supplier to raise awareness and give them the opportunity to investigate.Awareness training will be performed with the incoming inspection team to raise awareness.
 
Event or Problem Description
It was reported that the bd bd® whitacre spinal needle and bd¿ weiss epidural needle tray was cracked causing leakage.The following information was provided by the initial reporter: we are having major issues with the connectors for the catheter in our current cse kit.They crack and epidural solution is delivered to the bed instead of the epidural space.Please advise on next steps to correct the issue as this is a patient safety concern.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BD® WHITACRE SPINAL NEEDLE AND BD¿ WEISS EPIDURAL NEEDLE TRAY
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17456196
Report Number1625685-2023-00073
Device Sequence Number7153362
Product Code CAZ
UDI-Device Identifier00382904058289
UDI-Public(01)00382904058289
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number405828
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2023
Supplement Date Received by Manufacturer10/02/2023
Initial Report FDA Received Date08/03/2023
Supplement Report FDA Received Date10/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-