| Model Number |
MI1000 MED-EL CONCERT PIN |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
08/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user contacted audiology technical support after having a contraindicated mri done on (b)(6) 2023.The processor magnet reportedly repelled from the internal magnet and abnormal sound quality and sensation were experienced.The user sent to med-el the mri report that showed that a 3.0t mri was performed and there was no headwrap.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the information received from the field the recipient experienced magnet retention issue after magnetic resonance imaging (mri) performed at 3.0 tesla likely due to a demagnetization of the implant magnet.The instructions for use (ifu) for the concerned device clearly state that the concerned device is mri conditional for scanner field strengths of 0.2 tesla, 1.0 tesla and 1.5 tesla.Therefore the performed mri examination constitutes an abnormal use.Revision surgery is planned in (b)(6) 2023.
|
| |
|
Event or Problem Description
|
|
The user contacted audiology technical support after having a contraindicated mri done on (b)(6) 2023.She denied pain or discomfort during the mri but when she got home, she reported the processor magnet repelled from the internal magnet.When she tried to wear it, she experienced an abnormal sound quality and sensation.The user sent med-el the mri report that showed that a 3.0 t mri was performed and there was no head-wrap.The user has decided to go ahead with revision surgery, it is likely to take place in (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according to the information received from the field the recipient experienced magnet retention issue after magnetic resonance imaging (mri) performed at 3.0 tesla likely due to a demagnetization of the implant magnet.Device investigation confirmed a de-magnetized magnet.Other mechanical damages found are attributable to the removal surgery.The instructions for use (ifu) for the concerned device clearly state that the concerned device is mri conditional for scanner field strengths of 0.2 tesla, 1.0 tesla and 1.5 tesla.Therefore the performed mri examination constitutes an abnormal use.This is a final report.Correction to previous report: required intervention "yes".
|
| |
|
Event or Problem Description
|
|
The user contacted audiology technical support after having a contraindicated mri done on (b)(6) 2023.The user denied pain or discomfort during the mri but when she got home, she reported the processor magnet repelled from the internal magnet.When she tried to wear it, she experienced an abnormal sound quality and sensation.The user sent med-el the mri report that showed that a 3.0 t mri was performed and there was no head-wrap.The user has been reimplanted on (b)(6)2023.
|
| |
|
Event or Problem Description
|
|
The user contacted audiology technical support after having a contraindicated mri done on (b)(6) 2023.The user denied pain or discomfort during the mri but when she got home, she reported the processor magnet repelled from the internal magnet.When she tried to wear it, she experienced an abnormal sound quality and sensation.The user sent med-el the mri report that showed that a 3.0 t mri was performed and there was no head-wrap.The user has been reimplanted on (b)(6)2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according to the information received from the field the recipient experienced magnet retention issue after magnetic resonance imaging (mri) performed at 3.0 tesla likely due to a demagnetization of the implant magnet.Device investigation confirmed a de-magnetized magnet.Other mechanical damages found are attributable to the removal surgery.The instructions for use (ifu) for the concerned device clearly state that the concerned device is mri conditional for scanner field strengths of 0.2 tesla, 1.0 tesla and 1.5 tesla.Therefore the performed mri examination constitutes an abnormal use.This is a final report.
|
| |
|